Foil Gardon heat flow meter heat flow sensor

ДоговариваемыйОбновление на12/21
Модель
Природа производителя
Производители
Категория продукта
Место происхождения

Обзор

The foil type Gardon heat flux sensor GTT-25-227-WF meets the requirements of the national standard GB/T 25352-2010 (Test Method for Thermal Insulation and Sound Insulation Materials). To test the total heat flux density, the burner is calibrated to ensure an average heat flux of 18.2W/cm2. The nominal range of the foil type Gardon heat flux sensor GTT-25-227-WF is 0-22.7W/cm2. GTT-25-227-WF has the characteristics of good testing accuracy, high repeatability, durability, and easy use

Подробности о продукте

Our company is a hot flow manufacturer that supports customization!

Gardon foil heat flux sensor GTT-25-227-WF total heat flux density meter

GB/T 25352 standard heat flow meter


Foil Gardon heat flow meter heat flow sensorGTT-25-227-WF meets the national standardGB/T 25352-2010 (Test Method for Burnout Resistance of Thermal and Sound Insulation Materials).

The standard requirement is to calibrate the burner for testing the total heat flux density,Ensure that the average heat flux of the burner is 18.2W/cm2,Foil Gardon heat flux meterGTT-25-227-WF nominalThe range is 0-22.7W/cm2.

The foil Gardon heat flow meter GTT-25-227-WF has good testing accuracy,

High repeatability, sturdy and durable, easy to use and other characteristics.

Can be equipped with our DQ-100 data acquisition device to monitor heat flow in real-time

Calculate the average heat flux value.

The characteristics of the foil Gardon heat flow meter GTT-25-227-WF are:

Water cooling can ensure long-term continuous use;

The output is not affected by temperature, and the output signal is a mV level voltage;

(Can be converted into other signals with an additional converter)

Measure total heat flux

Accuracy better than 3%

Foil Gardon heat flux meterGTT-25-227-WF Technical Parameters:

Diameter: 25mm

Length: 25mm

Range: 22.7W/cm2

(also expressed as 227KW/m2)

Accuracy: ± 3%

Line length: 2m (also acceptable)lengthen, extend, elongate


The national standard GB/T 25352-2010 requires two types of heat flow meters, one for each type

Our model GTT-25-57-WF corresponds to a range of 5.7w/cm2.

Welcome to customization.

Our company is a manufacturer, and both GTT and STT series can be customized.